Verify load cell output, system capacity & resolution — Rice Lake 920i, Hardy HI 4050 / HI 6200 & more.
1 · Expected Load Cell Output
The #1 scale check: is the cell putting out the signal it should? Compare this to your meter.
920i exc 10 VHardy exc 5 V3 mV/V2 mV/V
Expected output @ load
—
Full-scale output
—
Implied load
—
⚠ Measured signal is above the cell's full-scale output — a dead giveaway of a damaged or bad load cell (or a wiring fault).
Output (mV) = Excitation × mV/V × Load ÷ Capacity. Healthy cells read ~0 at no load, climb linearly, and never exceed full-scale output. A signal above full-scale — or one that won't return to ~0 at no load — is a dead giveaway of a bad cell.
2 · System Capacity (summing box)
Multiple cells under a vessel. The empty vessel (dead load) eats part of your capacity.
Total system capacity
—
Net usable (live)
—
Dead load uses
—
Rated capacity per cell
—
Watch dead load — a heavy vessel on undersized cells wastes resolution.
3 · Display Resolution
Weight per division
—
Divisions = capacity ÷ increment. Keep it to what the system can actually resolve.
4 · Bridge Resistance Quick Reference
Disconnect the cell and ohm it out. Off-spec or low isolation = suspect cell.
Check
350 Ω cell
1000 Ω cell
Input (EXC+ ↔ EXC−)
~350–415 Ω
~1000–1100 Ω
Output (SIG+ ↔ SIG−)
~350 Ω
~1000 Ω
Any lead ↔ cell body
> 1000 MΩ (megger) — low = moisture/damage
Confirm exact values on the cell's spec sheet. Off-spec, or any reading to ground = failing cell.
5 · Load Cell Wiring Colors (Rice Lake)
Colors vary by maker AND model — always confirm against the cell's wiring guide. The two most common Rice Lake patterns:
Pattern B PC / RC / SB series, RLSB2, RLMK15, RLC…
+ Excitation
Green
Red
− Excitation
Black
Black
+ Signal
White
Green
− Signal
Red
White
+ Sense (6-wire)
Blue *
Brown *
− Sense (6-wire)
Brown *
Yellow *
Shield
Bare / Orange
Yellow
* Sense leads vary most by model (also seen Orange/Blue, Green/Grey). Colors are NOT universal across brands — when unsure, ohm it out and check the model's wiring sheet.
6 · Hardy C2 Load Cell Wiring
Hardy's standard C2 color code (HI 3000 / 4050 / 6200 / 6500 C2 load points). Non-C2 cells vary — check the supplier.
Wire
Color
+ Excitation (EXC+)
Red
− Excitation (EXC−)
Black
+ Signal (SIG+)
Green
− Signal (SIG−)
White
C2 + (data)
Grey
C2 − (data)
Violet
Shield
Ground / bare
C2+ (grey) & C2− (violet) are Hardy's electronic-calibration data lines — they let the module read the cell's capacity/output for C2 cal (no test weights) and per-cell diagnostics (Integrated Technician).
Some Hardy series use an "East Coast" code instead: +EXC Green, −EXC Black, +SIG White, −SIG Red (C2 still grey/violet). A standard C2 cable is these six wires plus shield — if a cell has extra wires, they're sense leads that tie to the excitation connections.
7 · Rice Lake 720i / 920i Connector (J1)
Load cell terminal pinout on the A/D card — land each wire (by its color, above) on the right pin.
Ties multiple load cells into one home-run cable to the indicator. Each cell lands on its own channel — same wire colors as above.
Hardy HI 6020IT / JB: up to 4 cells per box (8 with two boxes). The IT version reads each cell's voltage & weight individually — isolate a bad cell in seconds. Trim-pot version handles non-C2 cells. Rice Lake 4-channel: 4 cells, a trim pot per channel for corner balance.
Troubleshooting: uneven corner readings = a cell needs trim or is failing. On Hardy IT, switch to the internal test circuit to find the bad one.